Endomicroscope Controls

  • The tip of the endomicroscope is placed in gentle contact with the tissue
  • Laser light from the confocal scanning insert is projected and focussed to a fixed point
  • Buttons on the endoscope control body are used to move the position of the focal plane within the tissue
  • As the confocal scanning insert moves back within the endomicroscope tip, laser light is focussed to a point near the tissue surface
  • The confocal image plane is in a position near the surface of the tissue
  • When the confocal scanning insert moves forward within the endomicroscope tip, laser light is focussed to a point deeper in the tissue
  • The image plane is positioned deeper in the tissue
  • The pressure or position of the endomicroscope tip against the tissue remains unchanged when adjusting imaging depth

 

 